Tiêu chuẩn quốc tế

Số hiệu

Standard Number

ASTM F91-06(2019)
Năm ban hành 2019

Publication date

Tình trạng A - Còn hiệu lực

Status

Tên tiếng Anh

Title in English

Standard Practice for Testing for Leaks in the Filters Associated With Laminar Flow Clean Rooms and Clean Work Stations by Use of a Condensation Nuclei Detector
Chỉ số phân loại Quốc tế (ICS)

By field

13.040 - Chất lượng không khí
13.040.30 - Không khí vùng làm việc
Số trang

Page

2
Giá:

Price

Liên hệ / Contact us
Phạm vi áp dụng

Scope of standard

1.1 This practice covers the testing of the integrity of high-efficiency particulate air (HEPA) filters installed in laminar flow clean rooms of the ceiling to floor or wall to wall type, and laminar flow clean work stations. The recommended practice may be used to detect faults or voids in the filter media itself or in the joints between the filter and the room or work station structure. The determination of filter media efficiency is not within the scope of this practice. 1.2 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard. 1.2.1 Exception-The values given in parentheses in inch-pound units are for information only. 1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.